Table 4

Radiation exposure for screw implantation.

VariableGroup, mean ± SD (range) P
O-ArmSurgivisio
2D imaging duration, sec4 ± 3 (1–19)109 ± 45 (37–250) <0.01
No. of 3D acquisitions/patient1.1 ± 0.27 (1–2)1.6 ± 0.83 (1–5) <0.01
DLP, mGy.cm345 ± 161 (88–837)--
Overall DAP, mGy.cm2 -9664 ± 6978 (2000–29,276)-
DAP related to 2D, mGy.cm2 1167 ± 1111 (181–5645)4209 ± 4042 (406–25,365) <0.01
Overall E, mSv5.43 ± 2.53 (1.42–13.20)2.70 ± 2.32 (0.60–12.74) <0.01
E per level, mSv2.65 ± 1.46 (0.52–6.88)1.02 ± 0.87 (0.15–3.87) <0.01
E 2D, mSv0.26 ± 0.24 (0.04–1.41)1.16 ± 1.18 (0.12–6.88) <0.01
E 3D, mSv5.17 ± 2.42 (1.23–12.56)1.54 ± 1.47 (0.33–8.54) <0.01
E per 3D acquisition, mSv4.79 ± 2.03 (1.23–12.56)0.90 ± 1.06 (0.11–6.60) <0.01
E 2D calculated with w, mSv0.30 ± 0.29 (0.05–1.47)1.05 ± 1.06 (0.11–6.60) <0.01
AR, %2.7 × 10−4 ± 1.3 × 10−4 (6.4 × 10−5 to 6.7 × 10−4)1.4 × 10–4 ± 1.2 × 10−4 (3.0 × 10−5 to 6.3 × 10−4) <0.01
  • Abbreviations: AR, absolute risk of radiation-induced cancer; 2D, 2-dimensional; 3D, 3-dimensional; DAP, dose area product; DLP, dose length product; E, effective dose.

  • Note: Boldface indicates statistically significant findings.