Limitations of current in vitro test protocols for investigation of instrumented adjacent segment biomechanics: critical analysis of the literature

  • Review Article
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

Purpose

Accelerated degenerative changes at intervertebral levels adjacent to a spinal fusion, the so-called adjacent segment degeneration (ASD), have been reported in many clinical studies. Even though the pathogenesis of ASD is still widely unknown, biomechanical in vitro approaches have often been used to investigate the impact of spinal instrumentation on the adjacent segments. The goal of this review is (1) to summarize the results of these studies with respect to the applied protocol and loads and (2) to discuss if the assumptions made for the different protocols match the patients’ postoperative situation.

Methods

A systematic MEDLINE search was performed using the keywords “adjacent”, “in vitro” and “spine” in combination. This revealed a total of 247 articles of which 33 met the inclusion criteria. In addition, a mechanical model was developed to evaluate the effects of the current in vitro biomechanical test protocols on the changes in the adjacent segments resulting from different stiffnesses of the “treated” segment.

Results

The surgical treatments reported in biomechanical in vitro studies investigating ASD can be categorized into fusion procedures, total disc replacement (TDR), and dynamic implants. Three different test protocols (i.e. flexibility, stiffness, hybrid) with different loading scenarios (e.g. pure moment or eccentric load) are used in current biomechanical in vitro studies investigating ASD. According to the findings with the mechanical model, we found that the results for fusion procedures highly depend on the test protocol and method of load application, whereas for TDR and dynamic implants, most studies did not find significant changes in the adjacent segments, independent of which test protocol was used.

Conclusions

The three test protocols mainly differ in the assumption on the postoperative motion behavior of the patients, which is the main reason for the conflicting findings. However, the protocols have never been validated using in vivo kinematic data. In a parallel review on in vivo kinematics by Malakoutian et al., it was found that the assumption that the patients move exactly the same after fusion implemented with the stiffness- and hybrid protocol does not match the patients’ behavior. They showed that the motion of the whole lumbar spine rather tends to decrease in most studies, which could be predicted by the flexibility protocol. However, when the flexibility protocol is used with the “gold standard” pure moment, the difference in the kinematic changes between the cranial and caudal adjacent segment cannot be reproduced, putting the validity of current in vitro protocols into question.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Lund T, Oxland TR (2011) Adjacent level disk disease—is it really a fusion disease? Orthop Clin N Am 42:529–541

    Article  Google Scholar 

  2. Park P, Garton HJ, Gala VC et al (2004) Adjacent segment disease after lumbar or lumbosacral fusion: review of the literature. Spine (Phila Pa 1976) 29:1938–1944

    Article  Google Scholar 

  3. Hilibrand AS, Robbins M (2004) Adjacent segment degeneration and adjacent segment disease: the consequences of spinal fusion? Spine J 4:190S–194S

    Article  PubMed  Google Scholar 

  4. Malakoutian M, Volkheimer D, Street J et al (2015) Do in vivo kinematic studies provide insight into the degeneration of adjacent segments?—a qualitative systematic literature review. Eur Spine J. doi:10.1007/s00586-015-3992-0

    Google Scholar 

  5. Cole T-C, Burkhardt D, Ghosh P et al (1985) Effects of spinal fusion on the proteoglycans of the canine intervertebral disc. J Orthop Res 3:277–291

    Article  CAS  PubMed  Google Scholar 

  6. Bushell GR, Ghosh P, Taylor TKF et al (1978) The effect of spinal fusion on the collagen and proteoglycans of the canine intervertebral disc. J Surg Res 25:61–69

    Article  CAS  PubMed  Google Scholar 

  7. Hoogendoorn RJW, Helder MN, Wuisman PIJM et al (2008) Adjacent segment degeneration––observations in a goat spinal fusion study. Spine (Phila Pa 1976) 33:1337–1343

    Article  Google Scholar 

  8. Phillips FM, Reuben J, Wetzel FT (2002) Intervertebral disc degeneration adjacent to a lumbar fusion—an experimental rabbit model. J Bone Joint Surg Br 84-B:289–294

    Article  Google Scholar 

  9. Wilke H-J, Kettler A, Wenger KH, Claes LE (1997) Anatomy of the sheep spine and its comparison to the human spine. Anat Rec 247:542–555

    Article  CAS  PubMed  Google Scholar 

  10. Alini M, Eisenstein SM, Ito K et al (2008) Are animal models useful for studying human disc disorders/degeneration? Eur Spine J 17:2–19

    Article  PubMed Central  PubMed  Google Scholar 

  11. Reitmaier S, Schmidt H, Ihler R et al (2013) Preliminary investigations on intradiscal pressures during daily activities: an in vivo study using the merino sheep. PLoS One 8:e69610

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  12. Panjabi MM (2007) Hybrid multidirectional test method to evaluate spinal adjacent-level effects. Clin Biomech 22:257–265

    Article  Google Scholar 

  13. Panjabi MM (1988) Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine (Phila Pa 1976) 13:1129–1134

    Article  CAS  Google Scholar 

  14. Patwardhan AG, Havey RM, Meade KP et al (1999) A follower load increases the load-carrying capacity of the lumbar spine in compression. Spine (Phila Pa 1976) 24:1003–1009

    Article  CAS  Google Scholar 

  15. Wilke H-J, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  16. Gertzbein SD, Holtby R, Tile M et al (1984) Determination of a locus of instantaneous centers of rotation of the lumbar disc by Moiré fringes: a new technique. Spine (Phila Pa 1976) 9:409–413

    Article  CAS  Google Scholar 

  17. Quinnell RC, Stockdale HR (1981) Some experimental observations of the influence of a single lumbar floating fusion on the remaining lumbar spine. Spine (Phila Pa 1976) 6:263–267

    Article  CAS  Google Scholar 

  18. Dekutoski MB, Schendel MJ, Ogilvie JW et al (1994) Comparison of in vivo and in vitro adjacent segment motion after lumbar fusion. Spine (Phila Pa 1976) 19:1745–1751

    Article  CAS  Google Scholar 

  19. Lucas DB, Bresler B (1961) Stability of the ligamentous spine. Technical Report esr. 11 No. 40, Biomechanics Laboratory, University of California at San Francisco, The Laboratory

  20. Crisco JJ (1989) The biomechanical stability of the human lumbar spine: experimental and theoretical investigations (doctoral dissertation). CT, Yale University, New Haven

    Google Scholar 

  21. Pfeiffer M, Hoffman H, Goel VK et al (1997) In vitro testing of a new transpedicular stabilization technique. Eur Spine J 6:249–255

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  22. Bastian L, Lange U, Knop C et al (2001) Evaluation of the mobility of adjacent segments after posterior thoracolumbar fixation: a biomechanical study. Eur Spine J 10:295–300

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  23. Schmoelz W, Huber JF, Nydegger T et al (2003) Dynamic stabilization of the lumbar spine and its effects on adjacent segments: an in vitro experiment. J Spinal Disord Tech 16:418–423

    Article  CAS  PubMed  Google Scholar 

  24. Moore J, Yoganandan N, Pintar FA et al (2006) Tapered cages in anterior lumbar interbody fusion: biomechanics of segmental reactions. J Neurosurg Spine 5:330–335

    Article  PubMed  Google Scholar 

  25. Schmoelz W, Huber JF, Nydegger T et al (2006) Influence of a dynamic stabilisation system on load bearing of a bridged disc: an in vitro study of intradiscal pressure. Eur Spine J 15:1276–1285

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  26. Delank K-S, Gercek E, Kuhn S et al (2010) How does spinal canal decompression and dorsal stabilization affect segmental mobility? A biomechanical study. Arch Orthop Trauma Surg 130:285–292

    Article  PubMed  Google Scholar 

  27. Dahl MC, Ellingson AM, Mehta HP et al (2013) The biomechanics of a multilevel lumbar spine hybrid using nucleus replacement in conjunction with fusion. Spine J 13:175–183

    Article  PubMed  Google Scholar 

  28. Cheng BC, Gordon J, Cheng J, Welch WC (2007) Immediate biomechanical effects of lumbar posterior dynamic stabilization above a circumferential fusion. Spine (Phila Pa 1976) 32:2551–2557

    Article  Google Scholar 

  29. Akamaru T, Kawahara N, Yoon TS et al (2003) Adjacent segment motion after a simulated lumbar fusion in different sagittal alignments: a biomechanical analysis. Spine (Phila Pa 1976) 28:1560–1566

    Google Scholar 

  30. Umehara S, Zindrick MR, Patwardhan AG et al (2000) The biomechanical effect of postoperative hypolordosis in instrumented lumbar fusion on instrumented and adjacent spinal segments. Spine (Phila Pa 1976) 25:1617–1624

    Article  CAS  Google Scholar 

  31. Weinhoffer SL, Guyer RD, Herbert M, Griffith SL (1995) Intradiscal pressure measurements above an instrumented fusion: a cadaveric study. Spine (Phila Pa 1976) 20:526–531

    Article  CAS  Google Scholar 

  32. Chow DHK, Luk KDK, Evans JH, Leong JCY (1996) Effects of short anterior lumbar interbody fusion on biomechanics of neighboring unfused segments. Spine (Phila Pa 1976) 21:549–555

    Article  CAS  Google Scholar 

  33. Cunningham BW, Kotani Y, McNulty PS et al (1997) The effect of spinal destabilization and instrumentation on lumbar intradiscal pressure: an in vitro biomechanical analysis. Spine (Phila Pa 1976) 22:2655–2663

    Article  CAS  Google Scholar 

  34. Panjabi M, Malcolmson G, Teng E et al (2007) Hybrid testing of lumbar CHARITE discs versus fusions. Spine (Phila Pa 1976) 32:959–966

    Article  Google Scholar 

  35. Panjabi M, Henderson G, Abjornson C, Yue J (2007) Multidirectional testing of one- and two-level ProDisc-L versus simulated fusions. Spine (Phila Pa 1976) 32:1311–1319

    Article  Google Scholar 

  36. Panjabi MM, Henderson G, James Y, Timm JP (2007) StabilimaxNZ versus simulated fusion: evaluation of adjacent-level effects. Eur Spine J 16:2159–2165

    Article  PubMed Central  PubMed  Google Scholar 

  37. Molz FJ, Partin JI, Kirkpatrick JS (2003) The acute effects of posterior fusion instrumentation on kinematics and intradiscal pressure of the human lumbar spine. J Spinal Disord Tech 16:171–179

    Article  PubMed  Google Scholar 

  38. Strube P, Tohtz S, Hoff E et al (2010) Dynamic stabilization adjacent to single-level fusion: Part I. Biomechanical effects on lumbar spinal motion. Eur Spine J 19:2171–2180

    Article  PubMed Central  PubMed  Google Scholar 

  39. Dmitriev AE, Gill NW, Kuklo TR, Rosner MK (2008) Effect of multilevel lumbar disc arthroplasty on the operative- and adjacent-level kinematics and intradiscal pressures: an in vitro human cadaveric assessment. Spine J 8:918–925

    Article  PubMed  Google Scholar 

  40. Kikkawa J, Cunningham BW, Shirado O et al (2010) Biomechanical evaluation of a posterolateral lumbar disc arthroplasty device: an in vitro human cadaveric model. Spine (Phila Pa 1976) 35:1760–1768

    Article  Google Scholar 

  41. Ilharreborde B, Shaw MN, Berglund LJ et al (2011) Biomechanical evaluation of posterior lumbar dynamic stabilization: an in vitro comparison between Universal Clamp and Wallis systems. Eur Spine J 20:289–296

    Article  PubMed Central  PubMed  Google Scholar 

  42. Lindsey DP, Swanson KE, Fuchs P et al (2003) The effects of an interspinous implant on the kinematics of the instrumented and adjacent levels in the lumbar spine. Spine (Phila Pa 1976) 28:2192–2197

    Article  Google Scholar 

  43. Swanson KE, Lindsey DP, Hsu KY et al (2003) The effects of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976) 28:26–32

    Article  Google Scholar 

  44. Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA (2005) The effect of an interspinous process implant on facet loading during extension. Spine (Phila Pa 1976) 30:903–907

    Article  Google Scholar 

  45. Crawford NR, Brantley AGU, Dickman CA, Koeneman EJ (1995) An apparatus for applying pure nonconstraining moment to spine segments in vitro. Spine (Phila Pa 1976) 20:2097–2100

    Article  CAS  Google Scholar 

  46. Rohlmann A, Neller S, Bergmann G et al (2001) Effect of an internal fixator and a bone graft on intersegmental spinal motion and intradiscal pressure in the adjacent regions. Eur Spine J 10:301–308

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  47. Hartmann F, Dietz SO, Kuhn S et al (2011) Biomechanical comparison of an interspinous device and a rigid stabilization on lumbar adjacent segment range of motion. Acta Chir Orthop Traumatol Cech 78:404–409

    CAS  PubMed  Google Scholar 

  48. Cripton PA, Bruehlmann SB, Orr TE et al (2000) In vitro axial preload application during spine flexibility testing: towards reduced apparatus-related artefacts. J Biomech 33:1559–1568

    Article  CAS  PubMed  Google Scholar 

  49. Wilke H-J, Claes L, Schmitt H, Wolf S (1994) A universal spine tester for in vitro experiments with muscle force simulation. Eur Spine J 3:91–97

    Article  CAS  PubMed  Google Scholar 

  50. Cardoso MJ, Dmitriev AE, Helgeson M et al (2008) Does superior-segment facet violation or laminectomy destabilize the adjacent level in lumbar transpedicular fixation? An in vitro human cadaveric assessment. Spine (Phila Pa 1976) 33:2868–2873

    Article  Google Scholar 

  51. Tan J-S, Singh S, Zhu Q-A et al (2008) The effect of cement augmentation and extension of posterior instrumentation on stabilization and adjacent level effects in the elderly spine. Spine (Phila Pa 1976) 33:2728–2740

    Article  Google Scholar 

  52. Ingalhalikar AV, Reddy CG, Lim TH et al (2009) Effect of lumbar total disc arthroplasty on the segmental motion and intradiscal pressure at the adjacent level: an in vitro biomechanical study. J Neurosurg Spine 11:715–723

    Article  PubMed  Google Scholar 

  53. Cabello J, Cavanilles-Walker JM, Iborra M et al (2013) The protective role of dynamic stabilization on the adjacent disc to a rigid instrumented level. An in vitro biomechanical analysis. Arch Orthop Trauma Surg 133:443–448

    Article  CAS  PubMed  Google Scholar 

  54. Yoganandan N, Pintar F, Maiman DJ et al (1993) Kinematics of the lumbar spine following pedicle screw plate fixation. Spine (Phila Pa 1976) 18:504–512

    Article  CAS  Google Scholar 

  55. Ha S-K, Kim S-H, Kim DH et al (2009) Biomechanical study of lumbar spinal arthroplasty with a semi-constrained artificial disc (activ L) in the human cadaveric spine. J Korean Neurosurg Soc 45:169–175

    Article  PubMed Central  PubMed  Google Scholar 

Download references

Acknowledgments

We wish to thank the Alexander von Humboldt Foundation for their generous support of this research through a Research Award to Thomas R. Oxland during his sabbatical leave at the University of Ulm.

Conflict of interest

The authors affirm that they have no financial affiliation or involvement with any commercial organization that has direct financial interest in any matter included in this article.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hans-Joachim Wilke.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary material Table 1 Summary of the biomechanical articles addressing ALEs after spinal fusion (PDF 77.8 kb)

Supplementary material Table 2 Summary of the biomechanical articles addressing ALEs after TDR (PDF 28 kb)

586_2015_4040_MOESM3_ESM.pdf

Supplementary material Table 3 Summary of the biomechanical articles addressing ALEs after dynamic instrumentation (PDF 41.9 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Volkheimer, D., Malakoutian, M., Oxland, T.R. et al. Limitations of current in vitro test protocols for investigation of instrumented adjacent segment biomechanics: critical analysis of the literature. Eur Spine J 24, 1882–1892 (2015). https://doi.org/10.1007/s00586-015-4040-9

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00586-015-4040-9

Keywords

Navigation