Volume 23, Issue 9, November 2008, Pages 1095-1104
Clinical Biomechanics

Review
Biomechanical studies on cervical total disc arthroplasty: A literature review

https://doi.org/10.1016/j.clinbiomech.2008.06.002Get rights and content

Abstract

Background

Many models of cervical disc prostheses are currently commercially available or under clinical trial, and are based on several design concepts and built employing different materials. This paper is targeted to the understanding of the possible relationships between the geometrical, mechanical and material properties of the various cervical disc prostheses and the restoration of a correct biomechanics of the implanted spine.

Methods

Papers about cervical disc arthroplasty, based on ex vivo testing, mathematical models, and radiographic measurements, were included in the present review.

Findings

Although disc arthroplasty was found to be generally able to preserve a nearly physiological motion in the cervical spine, several alterations in the spine biomechanics due to disc arthroplasty were reported in the literature. An increase of the range of motion at the implanted level was observed in some ex vivo studies. Loss of mobility and heterotopic ossification were reported in radiographic investigations. Loss of lordosis at the implanted level was detected as well. Wear debris was usually found very limited and device stability seemed not to be an actual problem.

Interpretation

The possible relationships between the observed alterations in the spine biomechanics after disc arthroplasty and the properties of the various cervical disc prostheses have been reviewed. Clinical studies are needed to assess the validity of the considerations inferred from the biomechanical papers.

Introduction

Spinal fusion offers the surgeon an opportunity to remove the pathologic process, eliminate painful motion and obtain decompression of the neural elements (Hilibrand and Robbins, 2004). Fusion is at present the gold standard treatment for herniated cervical discs, and allows very high fusion rates, currently over 95% after application of anterior cervical implants (Mummaneni and Haid, 2004). However, several clinical studies show that fusion can be associated with symptomatic adjacent-segment disease (Goffin et al., 2003, Robertson et al., 2005), although the actual occurrence frequency of this complication remains an open question (Hilibrand and Robbins, 2004). Early adjacent level degeneration appears to be due to the abnormal kinematics and load transfer patterns at adjacent functional spinal units induced by fusion. Total disc arthroplasty (TDA), allowing the preservation of the mobility of the implanted segment, is believed to reduce the stress sustained by the adjacent levels and avoid, or at least slow down, their early degeneration (Albert and Eichenbaum, 2004). Clinical studies on TDA report no evidence of adjacent symptomatic disc degeneration, and no increased motion at adjacent levels after implantation of a disc prosthesis (Bertagnoli and Kumar, 2002, Duggal et al., 2004, McAfee et al., 2003).

Many models of disc prostheses are currently commercially available or under clinical trial, and are targeted to be used both in the cervical or in the lumbar spine. The prostheses generally consists in two or three components, mainly presenting articulating surfaces, and are built in advanced materials to ensure mechanical reliability and limit wear. The choice of the materials and the design of the shape of the device components influence the biomechanics of the implanted spine, thus affecting the clinical outcome. A number of biomechanical studies about TDA have been published, but the relationship between design and biomechanics of the implanted spine has not been clearly figured out yet. This is evidenced by the significant differences in the designs of the currently available devices.

This review summarizes the biomechanical studies concerning cervical TDA. Only single level TDA is considered, since multilevel and hybrid implantations induce significantly different biomechanical conditions which discussion would require a distinct review. The paper is structured along the targets of TDA which are related to biomechanical parameters:

  • restoration of a physiological kinematics and mobility, avoiding spinal instability;

  • protection of the biological structures, such as the adjacent intervertebral discs, the facet joints and the ligaments, from overloading;

  • restoration of a correct spinal alignment;

  • device stability and response to wear debris.

Aim of the present review is the understanding of the possible relationships between the aforementioned points and the geometrical, mechanical and material properties of the various currently available disc prostheses.

The papers available through a PUBMED (National Library of Medicine and National Institute of Health, USA) search (revised January 14th, 2008) were reviewed. Search strings employed were “cervical disc arthroplasty”, “cervical artificial disc”, “cervical disc replacement” and “cervical disc prosthesis”. Among the 309 papers found, 31 papers not written in the English language and 34 review papers were discarded. Furthermore, 190 papers were not considered, since they do not include any relevant biomechanical information on single level cervical TDA. Totally, 54 papers were reviewed.

Section snippets

Design concepts

Disc prostheses have been described, with reference to the allowed motion, using different classification systems (LeHuec et al., 2003, Link et al., 2004, Rousseau et al., 2006, Sekhon and Ball, 2005). In this review, the devices will be classified as unconstrained, semi-constrained and constrained, depending on the degrees of freedom allowed by the specific design (Fig. 1). All the three categories include prostheses featuring the widely employed “ball-and-socket” design, which consists in a

Finite axes of rotation

Most studies of the cervical spine have addressed flexion–extension, since these are believed to be the cardinal movements of this segment. The finite axes of rotation (FARs) of the cervical spine in flexion–extension have been quantitatively investigated in some studies (Amevo et al., 1991, Bogduk and Mercer, 2000, Penning and Wilmink, 1987). Generally, the FARs are located below the disc, posteriorly with respect to the endplate center (Fig. 2). From above downwards, the axis of rotation is

Protection of the biological structures

Although the most investigated issue on TDA biomechanics is motion preservation, the study of loads and stresses acting on the anatomical structures is gaining increasing attention. Indeed, unphysiological stresses in the tissues, which may be induced by the implantation of a disc prosthesis, are a direct cause of their early degeneration (Chang et al., 2007a, Hilibrand and Robbins, 2004). Some studies addressing the evaluation of stresses are currently available (Table 3). The two issues which

Spinal alignment

Disc arthroplasty can alter the spine alignment. In particular, resection of the anterior longitudinal ligament and the anterior portion of the annulus fibrosus, combined to an increase of the intervertebral space height, induces an alteration of the segmental lordosis (Cakir et al., 2005), regardless of the mechanics of the prosthesis itself.

In the same way as kinematics, spinal alignment after TDA is related to the design of the specific prosthesis. The use of the Bryan prosthesis can result

Device stability and wear

With reference to the interface between the disc prosthesis and the surrounding biological system, two topics are primarily discussed in the literature: the osseointegration at the bone–implant surface and the biological response to the possible formation of wear debris (Table 4). Besides the kinematics of the disc prosthesis, the materials and the possible presence of keels or screws are crucial with reference to these problems.

Most published animal studies report good or excellent

Conclusions

Based on the present literature review, TDA is generally able to preserve a nearly physiological motion in the cervical spine. However, several alterations in the cervical biomechanics due to TDA have been reported in the literature. Most authors reported RoMs nearly similar to the physiological values; nevertheless, RoM increase at the implanted level has been observed in some ex vivo studies. Loss of motion and heterotopic ossification have been reported in a significant number of patients.

Conflict of interest statement

All the authors (Fabio Galbusera, Chiara M. Bellini, Marco Brayda-Bruno, Maurizio Fornari) have no proprietary, financial, professional or other personal interest of any nature or kind in any product, service and/or company that could be construed as influencing the position presented in, or the review of, the present manuscript entitled “Biomechanical studies on cervical total disc arthroplasty: A literature review”.

References (73)

  • R.H. Bartels et al.

    Fusion around cervical disc prosthesis: case report

    Neurosurgery

    (2005)
  • R. Bertagnoli et al.

    Indications for full prosthetic disc arthroplasty: a correlation of clinical outcome against a variety of indications

    Eur. Spine J.

    (2002)
  • R. Bertagnoli et al.

    Early results after ProDisc-C cervical disc replacement

    J. Neurosurg. Spine

    (2005)
  • U.K. Chang et al.

    Range of motion change after cervical arthroplasty with ProDisc-C and prestige artificial discs compared with anterior cervical discectomy and fusion

    J. Neurosurg. Spine

    (2007)
  • U.K. Chang et al.

    Changes in adjacent-level disc pressure and facet joint force after cervical arthroplasty compared with cervical discectomy and fusion

    J. Neurosurg. Spine

    (2007)
  • D. Coric et al.

    Prospective randomized controlled study of the Bryan Cervical Disc: early clinical results from a single investigational site

    J. Neurosurg. Spine

    (2006)
  • J.C. Datta et al.

    Sagittal split fractures in multilevel cervical arthroplasty using a keeled prosthesis

    J. Spinal Disord. Tech.

    (2007)
  • de Jongh, C.U., Basson, A.H., Scheffer, C., 2007. Dynamic simulation of cervical spine following single-level cervical...
  • D.J. DiAngelo et al.

    An improved biomechanical testing protocol for evaluating spinal arthroplasty and motion preservation devices in a multilevel human cadaveric cervical model

    Neurosurg. Focus

    (2004)
  • D.J. DiAngelo et al.

    In vitro biomechanics of cervical disc arthroplasty with the ProDisc-C total disc implant

    Neurosurg. Focus

    (2004)
  • D.J. DiAngelo et al.

    Biomechanical testing of an artificial cervical joint and an anterior cervical plate

    J. Spinal Disord. Tech.

    (2003)
  • A.E. Dmitriev et al.

    Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty: an in vitro human cadaveric model

    Spine

    (2005)
  • N. Duggal et al.

    Early clinical and biomechanical results following cervical arthroplasty

    Neurosurg. Focus

    (2004)
  • N. Duggal et al.

    Traumatic loading of the Bryan Cervical Disc prosthesis: an in vitro study

    Neurosurgery

    (2007)
  • F. Galbusera et al.

    Biomechanics of the C5–C6 spinal unit before and after placement of a disc prosthesis

    Biomech. Model. Mechanobiol.

    (2006)
  • J. Goffin et al.

    Intermediate follow-up after treatment of degenerative disc disease with the Bryan Cervical Disc prosthesis: single-level and bi-level

    Spine

    (2003)
  • A.S. Hilibrand et al.

    Adjacent segment degeneration and adjacent segment disease: the consequences of spinal fusion

    Spine J.

    (2004)
  • N. Hu et al.

    Porous coated motion cervical disc replacement: a biomechanical, histomorphometric, and biologic wear analysis in a caprine model

    Spine

    (2006)
  • T. Ishii et al.

    Kinematics of the cervical spine in lateral bending: in vivo three-dimensional analysis

    Spine

    (2006)
  • T. Ishii et al.

    Kinematics of the subaxial cervical spine in rotation in vivo three-dimensional analysis

    Spine

    (2004)
  • W.K. Jensen et al.

    Bone ingrowth in retrieved Bryan Cervical Disc prostheses

    Spine

    (2005)
  • J.P. Johnson et al.

    Sagittal alignment and the Bryan cervical artificial disc

    Neurosurg. Focus

    (2004)
  • S.W. Kim et al.

    Effects of a cervical disc prosthesis on maintaining sagittal alignment of the functional spinal unit and overall sagittal balance of the cervical spine

    Eur. Spine J.

    (2008)
  • Y. Kotani et al.

    Multidirectional flexibility analysis of cervical artificial disc reconstruction: in vitro human cadaveric spine model

    J. Neurosurg. Spine

    (2005)
  • J.C. LeHuec et al.

    Shock absorption in lumbar disc prosthesis: a preliminary mechanical study

    J. Spinal Disord. Tech.

    (2003)
  • C. Leung et al.

    Clinical significance of heterotopic ossification in cervical disc replacement: a prospective multicenter clinical trial

    Neurosurgery

    (2005)
  • Cited by (73)

    • Comparative biomechanical analyses of lower cervical spine post anterior fusion versus intervertebral disc arthroplasty: A geometrically patient-specific poroelastic finite element investigation

      2022, Journal of Orthopaedic Translation
      Citation Excerpt :

      To decrease the risk of ASD and intersegmental immobility, total disc arthroplasty, or artificial disc replacement, has increasingly gained popularity in the last decade or so. Several types of artificial discs were patented based on metal-on-metal or metal-on-plastic designs, and few were successfully approved by the Food and Drug Administration (FDA) for the treatment of degenerative disc disease in the US and beyond [8,9]. Among these, the Bryan® cervical artificial disc (Medtronic, Memphis, TN, USA), classified within the metal-on-plastic design group, has demonstrated satisfactory clinical results [10,11], while the Prestige LP® cervical disc prosthesis (Medtronic, Memphis, TN, USA), considered as part of the newer generation of metal-on-metal devices, is known for allowing relatively simpler implantation techniques [11].

    • A novel bio-inspired hydrogel-based lattice structure to mechanically mimic human annulus fibrosus: A finite element study

      2021, International Journal of Mechanical Sciences
      Citation Excerpt :

      Intervertebral disc hernia and dysfunctions are the most common health problems that affect both young and old human bodies. Although the intervertebral disc replacements design and prosthetics production have known a remarkable evolution over recent years [1–5], discectomy and spinal fusion surgeries remain the most commonly preferred and used by surgeons [6–9]. Many generations of replacements have been developed.

    • Cervical Total Disc Replacement: Novel Devices

      2021, Neurosurgery Clinics of North America
      Citation Excerpt :

      CDA has also been shown to have lower reoperation and readmission rates over a 30-day period as well as over 5 years.7,8 A major reason for this is the ability to maintain or even restore motion to a diseased cervical spine segment to more closely resemble the native spine, without significantly affecting the adjacent levels.3,5,9–11 One of the principle concerns with regard to any spinal fusion procedure is the likelihood of affecting the adjacent segments.

    View all citing articles on Scopus
    View full text