The “slide technique”—a novel free-hand method of subaxial cervical pedicle screw placement

Abstract

Background

Cervical Pedicle Screw (CPS) placement is a challenging work due to the high risk of neurovascular complications. Although there have been several different free-hand or navigation assisted techniques for CPS placement, perforations may occur during screw insertion, especially lateral perforation. The objective of this manuscript is to describe a novel free-hand technique for subaxial CPS placement (C3–C7) and to evaluate if it decreases the chances of perforation.

Methods

Thirty-two patients undergoing surgery with CPS instrumentation (C3–C7) at our institute between June 2017 and December 2018 were included in this study. All the patients had cervical trauma, and pedicle screw insertion was performed according to the free-hand “slide technique”. The lamina, lateral mass and facet joint of the target area were exposed and the optimal entry point was found on the lateral mass posterior surface. A pedicular probe was then inserted and gently advanced. During the pedicle probe insertion, the cortex of the medial margin of the pedicle acted as a slide to permit the safe insertion of the screw. If the pedicle screw pathway was intact, the screw of the appropriate size was carefully placed. Three-dimensional (3D) CT imaging reconstruction was performed in all the patients after surgery, and screw perforations were graded with the Gertzbein-Robbins classification.

Results

Thirty-two patients who met the inclusion criteria were included in this study. A total of 257 CPSs (C3–7) were inserted, of which 41 CPSs were in C3, 61 CPSs were in C4, 55 CPSs were in C5, 53 CPSs were in C6, and 47 CPSs were in C7. The diameter and length of CPSs were 3.5 mm and 22–26 mm respectively. According to the Gertzbein-Robbins classification, grade 0, 231 screws; grade 1, 19 screws; and grade 2, 7 screws. No neurovascular complications occurred stemming from malpositioning of pedicle screws. Among perforated screws (26 screws), there were 16 lateral perforations, 5 medical perforations, and 4 inferior perforations.

Conclusions

The initial usage result shows the “slide technique” is a safe, effective and cost-effective technique for pedicle screw placement in the cervical spine. This is the first report of such a technique, and further studies are needed.

Peer Review reports

Background

Due to the excellent three-column stability, pedicle screws have been widely used in spinal surgeries, including spinal fracture [1, 2], deformity [3], tumors [4], etc. Although pedicle screws have been routinely used for lumbar and thoracic fixation, the safety of its usage in the cervical spine remains a concern. Instead, spinous process wire, lateral mass screws or facet screws rather than cervical pedicle screws (CPSs) are commonly utilized during cervical posterior fixation.

It is a fact that anatomical structure and the adjacent relationship of the cervical spine are complex and dangerous, while the cervical spinal cord is close to the medial pedicle cortex, and the vertebral artery is close to the lateral pedicle cortex. CPS placement is a very challenging work due to the high risk of neurovascular complications, including critical bleeding, cerebral infarction, paralysis, and even death. Since Abumi et al. [5] first described the CPS placement method in 1994, many researchers have proposed various improved methods for this operation [6,7,8,9,10]. However, there is no ideal solution which can make CPS to be routinely used for cervical posterior fixation up to now.

With the advancement of technology in the field of digital navigation, the safety, as well as the accuracy of CPS placement, have been improved greatly [11]. However, the navigation system has its limitations, such as high cost [12], tedious procedures, and relatively limited indications [12].

Inspired by the “slide technique” in the thoracic spine [13], we devised a novel method for free-hand CPSs insertion also called “slide technique”, to increase the accuracy of CPS placement via a direct slide on the cortex of the medial margin of the pedicle. The purpose of this study is to describe the “slide technique” in CPS placement and to evaluate if it decreases the chances of perforation .

Methods

Study design

This was a clinical prospective study from June 2017 to December 2018. All the data were analyzed anonymously. An informed consent waiver was granted. This study was approved by the medical ethics committees of Hunan Provincial People’s Hospital (NO.2017-S65).

Patients

Patients with cervical trauma underwent posterior cervical surgeries and the CPS instrumentation aimed at reconstructing cervical stability were included. Patients with cervical pedicle stenosis, dysplasia or severe destruction that cannot insert CPS were excluded.

A total of 32 patients who met the inclusion criteria were included in this study. All patients came from the department of spinal surgery, Hunan Provincial People’s Hospital. Their mean age was 52.8 years (range, 28–74 years). There were 23 male and 9 female patients.

Surgical technique

All patients underwent a Three-dimensional (3D) Computed Tomography (CT) scan of the cervical spine to measure pedicle diameter and appropriate screw length before the operation.

All patients were placed in the prone position with the head fixed using brachiocephalic traction. After successful general anesthesia, we made a standard midline skin incision on the posterior neck. Subperiosteal dissection was performed along the spinous process and lamina. Lateral mass and facet joint of the target area were exposed. The target vertebral lateral mass was divided into three equal parts by longitudinal lines and the optimal entry point was chosen at a point on the lateral longitudinal line and slightly below the inferior margin of the facet joint. The distance between the inferior margin of the facet joint and entry point needed to be measured on X-ray or CT images before operation (Fig. 1). First, the cortex at the entry point was penetrated with a high-speed burr or awl. The pedicle probe was blunt-tipped with bent head, and its diameter was 2.1 mm(Fig. 2). After entered from the entry point with head bend inside, the pedicle probe rotated while gently advancing at a maximum angle on more than 45° medially. Probe stopped when meeting resistance. At this time, the probe head had reached the cortex of the medial margin of the pedicle (Fig. 3a). Thereafter, by rotating the probe by 180 (its head became to bend outward) (Fig. 3b) and reducing the medial angle, the pedicle probe avoided the medial cortex easily and it could continue to advance to the vertebrae body. During this procedure, the cortex of the medial margin of the pedicle acted as a slide to permit the safe insertion of the probe (Fig. 3c). Once the probe reached a depth of 15 mm (head of probe had reached the body), it was rotated by 180 ° again to make its head bend inside (Fig. 3d). After increasing the medial angle, the pedicle probe then continued to insert and gently advanced to a depth of 20 mm (Fig. 3e). We rotated the probe by 360° twice to expand the pathway (Fig. 3f). A ball-tip probe was used to evaluate the integrity of the pedicle screw pathway. If the pathway was intact, the screw of the appropriate size was carefully placed (Fig. 3g). The blue line in Fig. 3h represented the changes of the pathway during the probing, and the red line represented the final screw pathway. C-arm fluoroscopy was used to preliminarily evaluate the screw position during operation, especially to avoid superior and inferior perforations.

Fig. 1
figure 1

Optimal entry point (frontal view)

Fig. 2
figure 2

Diagram of the pedicle probe

Fig. 3
figure 3

Operative technique (axial view)

The key point of the “slide technique” was to use the cortex of the medial margin of the pedicle as a “slide” to permit correct probe positioning. The theoretical supports of this novel method includes 1. the cortex of the medial margin of the cervical pedicle was 1.4–3.6 times thicker than that of the lateral margin [14], so it was not easy to break through when negotiating, tapping and placing a screw. In comparison, the cortex of the lateral margin was thin and easy to be perforate. Previous studies had confirmed that the incidence rate of lateral perforation was significantly higher than that of the medial perforation during CPS insertion [15, 16]. 2. There were buffer tissues such as dural sac and epidural fat between the medial pedicle cortex and cervical spinal cord. Cases of medical perforation without spinal cord injury often could be seen in clinical work. In contrast, the vertebral artery closed to the lateral pedicle cortex was easy to be damaged due to the limitation of the vertebroarterial foramen. Therefore, we believed that it was safe and reliable to select the cortex of the medial margin of the pedicle for sliding.

Evaluation of security and accuracy

3D-CT imaging reconstruction was performed in all the patients after cervical surgeries, to assess pedicle screw instrumentation. We observed whether CPSs penetrate the pedicle cortex or not and if so, measured the distance between screws and the pedicle cortex. According to Gertzbein-Robbins classification [17], the degree of perforation was classified into four grades on postoperative CT scans. If the CPS located within the pedicle and did not perforated, that was defined as grade 0. Grade 1 perforation was defined if perforation was less than 2 mm. If perforation was 2–4 mm, it was classified as grade 2. Perforation of more than 4 mm was defined as grade 3. We also recorded the direction of perforation, including lateral perforation, medical perforation, superior perforation, and inferior perforation.

Besides, we reviewed the clinical information of all patients. Complications directly related to CPSs placement were recorded in all patients, such as spinal cord injury, nerve root injury, vertebral artery injury/rupture, aortic injury, etc.

The sample size of our study was calculated from our clinical experience (α error, 0.05; power, 0.8; P 0.9; P0, 0.8). The required sample size was at least 108 CPSs. We used SPSS Version 22.0 statistical software for all analyses.

Results

A total of 257 CPSs (C3–7) were inserted, of which 41 CPSs were in C3, 61 CPSs were in C4, 55 CPSs were in C5, 53 CPSs were in C6, and 47 CPSs were in C7. The diameter and length of CPSs were 3.5 mm and 22–26 mm respectively. According to the Gertzbein-Robbins classification, 231 screws (89.9%) were grade 0; 19 screws (7.4%) were grade 1; and 7 screws (2.7%) were grade 2. In perforated screws (26 screws), lateral perforations were 16(61.5%), medical perforations were 5(19.2%), and inferior perforations were 4(15.4%).

No neurovascular complications occurred stemming from malpositioning of a pedicle screw during operation. None of these patients showed neurological deterioration after the surgery or symptoms related to vertebral artery injury/rupture. There was also no nerve root injury happened after CPSs placement. A typical case is shown in Fig. 4.

Fig. 4
figure 4

Example of a patient with cervical trauma. a Preoperative cervical CT scan, b Preoperative cervical MR imaging, c Sagittal CT image after cervical operation, d-i transverse CT images after cervical operation

Discussion

Cervical fixation using pedicle screws is becoming increasingly popular for various cervical diseases due to their advantages, such as three-column fixation, sagittal correction, and excellent biomechanical stability when compared with spinous process wire, lateral mass screws or facet screws [18, 19]. However, the CPS has not become the routine for posterior cervical fixation due to the high risk and high skills required.

In addition to the CPS, lateral mass screw (LMS) is another commonly used method for posterior subaxial cervical fixation. Previous researches and clinical practices showed that these two methods have their advantages and disadvantages [19]. The placement of LMS is comparatively simple, and the incidence of neurovascular complications during operation is lower [20]. However, LMS showed a small amount of bony purchase and thereby lesser pullout strength especially in the presence of osteoporosis in biomechanical experiments when compared with CPS. By contrast, the biomechanical result showed the pullout strength was significantly higher for the CPS, and thus it was relatively difficult to loosen [19]. The technique for CPS placement is more difficult than LMS although much successful clinical use of CPSs has been described. Reports of the CPS insertion failure have shown the possibility of serious iatrogenic injury [16, 21].

The most challenging process of placing CPSs safely into the cervical vertebrae is making the exact cancellous pathway between the vertebral artery and cervical spinal cord for the pedicle screw. Lateral pedicle perforation when placing CPSs may injury the vertebral artery, while medial pedicle perforations can injury the cervical spinal cord.

The cortex of the medial margin of the cervical pedicle is much thicker and stronger than that of the lateral margin, so it is predisposed to lateral perforation during pathway preparation, tapping, and insertion of the screws. Previous literature has confirmed this point of view. A multicenter study from Japan on the complications of CPS placing during the conventional free-hand technique showed that 75% (57/76) of all misplaced screws were lateral pedicle perforation, while only 25% (19/76) were medial pedicle perforation [16]. Therefore, attention needs to be paid to how to reduce lateral pedicle perforation while designing a new CPS placing technology.

The most representative conventional free-hand technique for subaxial CPS placing is the method proposed by Abumi et al. [5] in 1994. The contents include: the entry point at the posterior cortex of the articular mass was determined slightly lateral to the center of the articular mass and close to the posterior margin of the superior articular surface. The intended angle of the screws based on measurements of preoperative CT images was 30–40° medial to the midline in the transverse plane, and parallel to the upper end-plate in the sagittal plane. The insertion of the screw was greater than two-thirds of the AP vertebral body depth. Subsequently, Jeanneret [6], Miller [7], and Liu [8] proposed various improved free-hand methods for CPS placement. There were also studies reported the CPS placement using the medial funnel technique [22] and medial cortical pedicle screw technique [9]. Burcev et al. [10] introduced a standardized and fast method for subaxial CPS: screw insertion based on the simple angles to the bony landmarks. However, the accuracy of conventional CPS placement methods needs to be improved though there are many methods to choose from.

In recent years, digital navigation technology, 3D printing technology, and robotic technology are also arising in CPS placement. For example, Ishikawa et al. [23] placed 108 cervical CPSs using an intraoperative, full-rotation, 3D image (O-arm)-based navigation system. The results showed that 96 of them (88.9%) were grade 0, 9 were grade 1 and 3 were grade 2. There were no complications such as vascular and nervous complications, indicating that a combination of intraoperative 3D image-based navigation with other techniques may result in more accurate CPS placement. But not all the results about using the navigation in CPSs placement were optimistic. Uehara et al. [15] inserted the CPSs by using a CT-based navigation system during operations. The results showed that the combined rate of grades 2 and 3 perforations was 20.0% (116/579). Therefore, the authors concluded careful insertion of pedicle screws is necessary, especially at C3 to C5, even when using a CT-based navigation system. There were a few studies even showing more perforations with navigated screws than with the free-hand technique [24, 25]. Although most of the studies support that navigation technology can improve the accuracy of pedicle screw placement [11], none of the high-tech can completely avoid the occurrence of screw perforations. The navigation system and other computer-aided technology always associated with complex operation procedures and expensive equipment costs [12]. Moreover, due to the high mobility of the cervical spine, cervical spine alignment can easily change during operation [26], which may lead to inaccurate synchronization to preoperative images. All the shortcomings limit the application of digital navigation technology, especially in developing countries.

Therefore, it is of great clinical significance to develop a cost-effective and safe method for free-hand CPS placement. Raphael Vialle et al. [13] developed the “sliding technique” for safe pedicle screw placement in the thoracic spine in 2004. The key point of this novel technique was to use the cortex of the anterior aspect of the transverse process and the lateral margin of the pedicle as a “slide” to permit correct probe positioning during pedicle probe insertion. Inspired by it, we devised a novel method for free-hand CPSs insertion also called “slide technique”, to increase the accuracy of CPS placement via a direct slide on the cortex of the medial margin of the pedicle.

Preliminary clinical results of this novel technique showed a higher rate of correct screw position, comparable to CPS placement with navigation system. Meanwhile, no neurovascular complications occurred stemming from malpositioning of pedicle screws. In the process of this free-hand technique, screw perforations inevitably occurred, with lateral pedicle perforation accounted for the majority, which was consistent with previous studies.

Several points should be paid attention to in the clinical application of the “sliding technique”. First, it is necessary to carefully read and analyze the imaging data before an operation, especially to accurately measure the diameter of the pedicle and pay attention to the variation of anatomical structure (such as pedicle sclerosis, pedicle slenderness, vertebral foramen malformation, local bone destruction, etc.). Second, the pedicle probe should be rotated while gently being advanced, and be stopped in case of resistance. No violence should be used during the whole operation. To ensure the accuracy of the pathway, intraoperative fluoroscopy is necessary. Third, when CPS placement is difficult, lateral mass screw or other fixation may be selected.

Please note that the patients included in the current study were all with cervical trauma, whose pedicle variation rate was small, so the accuracy of screw placement was relatively easy to achieve. The accuracy of the screw may decrease when applied to difficult situations such as cervical deformity.

Conclusions

To our knowledge, this is the first report of slide technique use in CPS placement. This novel free-hand “slide technique” could be considered as a safe, effective and cost-effective method for CPS placement.

However, this study is also limited due to the relatively small number of screws used and uniform criterion for pedicle perforation. Besides, all the CPSs placements were evaluated by the researcher involved in the surgery, which could have led to bias. A multicenter large sample study is desired to establish in the future.

Availability of data and materials

The datasets used and analysed in the current study are available from the corresponding author on reasonable request.

Abbreviations

CPS:

Cervical Pedicle Screw

3D:

Three-Dimensional

CT:

Computed Tomography

LMS:

Lateral Mass Screw

References

  1. Sjöström L, Karlström G, Pech P, Rauschning W. Indirect Spinal Canal Decompression in Burst Fractures Treated With Pedicle Screw Instrumentation. Spine. 1996;21(1):113–23.

  2. Carl AL, Tromanhauser SG, Roger DJ. Pedicle Screw Instrumentation for Thoracolumbar Burst Fractures and Fracture-Dislocations. Spine. 1992;17(8):317–24.

  3. Kwan MK, Chiu CK, Gani SMA, CCY W. Accuracy and Safety of Pedicle Screw Placement in Adolescent Idiopathic Scoliosis Patients: A Review of 2020 Screws using computed tomography assessment. Spine. 2017;42(5):326–35.

  4. Oda I, Abumi K, Ito M, Kotani Y, Oya T, Hasegawa K, et al. Palliative Spinal Reconstruction Using Cervical Pedicle Screws for Metastatic Lesions of the Spine. Spine. 2006;31(13):1439–44.

  5. Abumi K, Itoh H, Taneichi H, Kaneda K. Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord. 1994;7(1):19–28.

    Article  CAS  Google Scholar 

  6. Jeanneret B, Gebhard JS, Magerl F. Transpedicular Screw Fixation of Articular Mass Fracture-Separation. J Spinal Disord. 1994;7(3):222–9.

  7. Miller RM, Ebraheim NA, Xu R, Yeasting RA. Anatomic Consideration of Transpedicular Screw Placement in the Cervical Spine. Spine. 1996;21(20):2317–22.

  8. Liu J, Li Y, Wu Y, Zhu Q. A Novel Method of Cervical Pedicle Screw Placement From C3 to C5 and Its Clinical Applications. Spine. 38(8):E504–E12.

  9. Mahesh B, Upendra B, Mahan RS. The medial cortical pedicle screw—a new technique for cervical pedicle screw placement with partial drilling of medial cortex. Spine J. 2014;14(2):371–80. .

  10. Burcev A, Pavlova O, Diachkov K, Diachkova G, Gubin A. Easy method to simplify “freehand” subaxial cervical pedicle screw insertion. J Craniovertebral Junction Spine. 2017;8(4):390.

    Article  Google Scholar 

  11. Luther N, Iorgulescu JB, Geannette C, Gebhard H, Saleh T, Tsiouris AJ, et al. Comparison of navigated versus non-navigated pedicle screw placement in 260 patients and 1434 screws. J Spinal Disord Tech. 2015;28(5):E298.

    Article  Google Scholar 

  12. Yukawa Y, Kato F, Ito K, Horie Y, Hida T, Nakashima H, et al. Placement and complications of cervical pedicle screws in 144 cervical trauma patients using pedicle axis view techniques by fluoroscope. Eur Spine J. 2009;18(9):1293–9.

    Article  Google Scholar 

  13. Vialle R, Zeller R, Gaines RW. The “slide technique”: an improvement on the “funnel technique” for safe pedicle screw placement in the thoracic spine. Eur Spine J. 2014;23(4):452–6.

    Article  Google Scholar 

  14. Panjabi M, Shin E, Chen NC, Wang JL. Internal morphology of human cervical pedicles. Spine. 2000;25(10):1197–205.

    Article  CAS  Google Scholar 

  15. Uehara M, Takahashi J, Ikegami S, Mukaiyama K, Kuraishi S, Shimizu M, et al. Screw perforation features in 129 consecutive patients performed computer-guided cervical pedicle screw insertion. Eur Spine J. 2014;23(10):2189–95.

    Article  Google Scholar 

  16. Hiroaki N, Yasutsugu Y, Shiro I, Tokumi K, Mitsuhiro K, Makoto Y, et al. Complications of cervical pedicle screw fixation for nontraumatic lesions: a multicenter study of 84 patients. J Neurosurg Spine. 2012;16(3):238–47.

    Article  Google Scholar 

  17. Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo. Spine. 1990;15(1):11–4.

    Article  CAS  Google Scholar 

  18. Kotani Y, Cunningham BW, Abumi K, Mcafee PC. Biomechanical Analysis of Cervical Stabilization Systems. Spine. 1994;19(Supplement):2529–39.

  19. Jones EL, Heller JG, Silcox DH, Hutton WC. Cervical Pedicle Screws Versus Lateral Mass Screws. Spine. 1997;22(9):977–82.

  20. Yoshihara H, Passias PG, Errico TJ. Screw-related complications in the subaxial cervical spine with the use of lateral mass versus cervical pedicle screws. J Neurosurg Spine. 19(5):614–23.

  21. Zhang J, Xu R, Li Z, Zha W. Cerebral infarction due to malposition of cervical pedicle screw: A case report. Medicine(Baltimore). 2018;97(7):e9937.

  22. Lee JH, Choi BK, Han IH, Choi WG, Nam KH, Kim HS. Cervical pedicle screw placement using medial funnel technique. Korean J Spine. 2017;14(3):84–8.

    Article  Google Scholar 

  23. Ishikawa Y, Kanemura T, Yoshida G, Matsumoto A, Ito Z, Tauchi R, et al. Intraoperative, full-rotation, three-dimensional image (O-arm)-based navigation system for cervical pedicle screw insertion. J Neurosurg Spine. 2011;15(5):472–8.

    Article  Google Scholar 

  24. Ludwig SC, Kowalski JM, Edwards CC, Heller JG. Cervical pedicle screws: comparative accuracy of two insertion techniques. Spine. 2000;25(20):2675–81.

    Article  CAS  Google Scholar 

  25. Kast E, ., Mohr K, ., H-P R, W Börm. Complications of transpedicular screw fixation in the cervical spine. Eur Spine J 2006;15(3):327–334.

    Article  CAS  Google Scholar 

  26. Yoshii T, Hirai T, Sakai K, Inose H, Kato T, Okawa A. Cervical pedicle screw placement using intraoperative computed tomography imaging with a mobile scanner gantry. Eur Spine J. 2016;25(6):1690–7.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank the patients, physicians, and administrative staff who made this study possible.

Funding

This work is supported by Scientific research project of Hunan Health Committee (NO. 20200042).

Author information

Authors and Affiliations

Authors

Contributions

LB conceived and designed this study, performed the literature searches, extracted the data, interpreted the data and drafted the manuscript. LXY, SXJ, and WGP participated in performing the literature searches and drafting the manuscript. CYX revised the manuscript and acted as guarantor for the paper. The guarantors accept full responsibility for the conduct of the study, had access to the data, and enrolled the decision to publish. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Yixin Chen.

Ethics declarations

Ethics approval and consent to participate

All patients receive written information among their participation in the study. They sign a written consent to participate in the study. The Ethics Committee of Hunan Provincial People’s Hospital approved current study (NO.2017-S65). All clinical investigations were performed in accordance with the guidelines of the Declaration of Helsinki.

Consent for publication

Not Applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Liu, B., Liu, X., Shen, X. et al. The “slide technique”—a novel free-hand method of subaxial cervical pedicle screw placement. BMC Musculoskelet Disord 21, 399 (2020). https://doi.org/10.1186/s12891-020-03420-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-020-03420-0

Keywords