Asian Spine J Search

CLOSE


Asian Spine J > Volume 13(6); 2019 > Article
Kam, Gan, Dimou, Awad, Kavar, Nair, and Morokoff: Learning Curve for Robot-Assisted Percutaneous Pedicle Screw Placement in Thoracolumbar Surgery

Abstract

Study Design

Retrospective review of an initial cohort of consecutive patients undergoing robot-assisted pedicle screw placement.

Purpose

We aimed to evaluate the learning curve, if any, of this new technology over the course of our experience.

Overview of Literature

Percutaneous pedicle screws have specific advantages over open freehand screws. However, they require intraoperative imaging for their placement (e.g., fluoroscopy and navigation) and require increased surgeon training and skill with the learning curve estimated at approximately 20–30 cases. To our knowledge, this is the first study that measures the learning curve of robot-guided purely percutaneous pedicle screw placement with comprehensive objective postoperative computed tomography (CT) scoring, time per screw placement, and fluoroscopy time.

Methods

We included the first 80 consecutive patients undergoing robot-assisted spinal surgery at Melbourne Private Hospital. Data were collected for pedicle screw placement accuracy, placement time, fluoroscopy time, and revision rate. Patient demographic and relevant perioperative and procedural data were also collected. The patients were divided equally into four sub-groups as per their chronological date of surgery to evaluate how the learning curve affected screw placement outcomes.

Results

Total 80 patients were included; 73 (91%) had complete data and postoperative CT imaging that could help assess that placement of 352 thoracolumbar pedicle screws. The rate of clinically acceptable screw placement was high (96.6%, 95.4%, 95.6%, and 90.7%, in groups 1 to 4, respectively, p=0.314) over time. The median time per screw was 7.0 minutes (6.5, 7.0, 6.0, and 6.0 minutes in groups 1 to 4, respectively, p=0.605). Intraoperative revision occurred in only 1 of the 352 screws (0.3%).

Conclusions

We found that robot-assisted screw placement had high accuracy, low placement time, low fluoroscopy time, and a low complication rate. However, there were no significant differences in these parameters at the initial experience and the practiced, experience placement (after approximately 1 year), indicating that robot-assisted pedicle screw placement has a very short (almost no) learning curve.

Introduction

Percutaneous pedicle screws have become the standard of care in spinal fusion surgery because of their multiple potential advantages, including more lateral placement, avoidance of the superior facet joint, stronger biome-chanical pull out strength, smaller incisions, less bleeding, lower pain, less infection, and possibly better long-term outcomes [1-8]. However, they depend on some form of intraoperative imaging for placement (e.g., fluoroscopy and computer guided image navigation) and require high-level surgeon training and skill, with the learning curve being completed, in terms of operative time and complications, once the surgeon has performed about 20–30 screw placements [9,10].
Robot-assisted pedicle screw placement ensures anatomical orientation and mechanical control of the instrumentation (placement of the cannula, drill position and trajectory, and placement of the pedicle screw). This provides additional advantages over navigated or fluoroscopy-guided pedicle screw placement by reducing the 6° of motion in a three-dimensional space of Jamshidi needle placement to 2° of freedom (depth of tool insertion and tool clockwise/counter-clockwise rotation). Moreover, it only uses 2 shots of bi-planar radiography to register to the preoperative computed tomography (CT); therefore, the Renaissance robot can significantly save fluoroscopy time and staff radiation exposure.
This is a relatively new technology; therefore, limited studies have been conducted on the accuracy and advantages of Renaissance robot. Currently, there is growing evidence that robot-guided screws are more accurate than conventional methods of percutaneous pedicle screw placement with most retrospective studies and trials reporting nearly 99% accuracy of clinically acceptable screw positioning [11-16].
We aimed to evaluate the learning curve, if any, of this new technology by measuring the screw placement time and accuracy, fluoroscopy exposure, revision rate, and complications over the course of our experience. To our knowledge, this is the first study to assess the learning curve of robot-guided purely percutaneous pedicle screw placement with comprehensive objective postoperative CT scoring, time per screw, and fluoroscopy time.

Materials and Methods

1. Study setting and participants

Between February 2016 and May 2017, when robot-assisted pedicle screw placement was first instituted, 80 consecutive patients for whom spinal surgery was performed by the three senior authors (AM, BK, and GN) were included. The surgeries were performed for several indications; the most common indication being degenerative disk disease, at Melbourne Private Hospital. All the surgeons have a subspecialty spinal practice as neurosurgeons. The results were obtained from retrospective records and perioperative imaging review. All the cases involved percutaneous pedicle screw placement. Individual pedicle screw placement accuracy, time, fluoroscopy time, and revision rate were recorded. Patient demographic and relevant data, such as surgical indication, comorbidities, number of spinal levels undergoing surgery, and complications were recorded. Human Research Ethics Committee (HREC 2017.202) approval was obtained for the study, and written consent was not required for retrospective data review. All patients undergoing our initial robotassisted thoracolumbar surgery experience were included; therefore, there were no specific exclusion criteria. The patients were divided equally into four sub-groups as per their chronological date of surgery to evaluate how the learning curve affected screw placement accuracy, fluoroscopy time, and screw placement time. Patients were excluded if incomplete data were available.

2. Outcomes

For each case, screw placement accuracy, screw placement time, fluoroscopy time, and revision rate were measured prospectively. Each screw was scored independently by two authors (JK and CG) using the Gertzbein-Robbins classification [17] (Fig. 1) based on the postoperative CT images obtained for all patients [18,19]. Any discrepancy was resolved by consensus and review. Complications and adverse events were also reviewed. The outcomes were compared between the groups.

3. Surgical technique

All the patients underwent percutaneous screw placement using The Renaissance Guidance System (Mazor Robotics Ltd., Caesarea, Israel; ‘Mazor robot’), a navigation device used in spinal surgery for the placement of screws into the spine. The technique involves matching an on-table bi-planar fluoroscopic image to a preoperative CT of the spine, wherein pre-planned screws have been designed on a computer by the surgeon. A mechanical robot arm attached to the table and/or patient then moves to place a guide tube that assists the surgeon to drill a pilot hole and accurately place a guide wire over which a spinal screw can be inserted.

4. Statistical analyses

Group comparisons were conducted using Fisher’s exact, chi-square, and Freeman-Halton extension tests with a significance level of p<0.05. Continuous variables were compared using the Mann-Whitney U, Kruskal-Wallis, and one way analysis of variance tests, as appropriate. IBM SPSS ver. 20.0 (IBM Corp., Armonk, NY, USA) was used to perform the statistical analyses and generate graphs.

Results

1. Patients

Total 80 patients were admitted to undergo robot-assisted spinal surgery for several indications within the study period (Fig. 2). Baseline characteristics of the groups are described in Table 1. Complete data on demographic profile, presentation details, complications, and postoperative CT imaging to assess hardware placement were available for 73 patients (91%). The mean±standard deviation age of the study participants was 66.6±11.1 years, and 63.0% were women. There were no significant differences between the groups for comorbidities, number of levels operated, indication, or approach. Significantly higher numbers of patients underwent thoracic surgery in the third quarter of the group’s experience (Table 1).

2. Primary outcomes

Total 352 thoracolumbar percutaneous pedicle screws were placed for 73 patients; of these, 90.1% were placed in an ideal position (grade A) without evidence of breach. Distributions of ideal versus non-ideal placement of pedicle screws are given in Table 2. The rate of ideal screw placement remained unchanged among the groups. The average (median) time for screw placement was 7 minutes and did not vary significantly among the groups. Fig. 3 demonstrates screw placement time over the authors’ experience that remained low from the beginning of the series. The average (median) fluoroscopy time per screw was 3 seconds. This varied significantly among the groups, with group 3 (third quarter) having a significantly (p=0.021) increased fluoroscopy time per screw of 4 seconds.

3. Secondary outcomes and adverse events

There were no misplaced screws placed within the study period. There was one intraoperative revision (0.3%) that was not associated with complication (Table 3). Overall, there were four complications (5.5%). One patient experienced a spinal cord stroke associated with neurological deficit. One patient underwent revision surgery for an asymptomatic extruded bone graft detected on postoperative imaging. One patient had two freehand pedicle screws placed after dysfunction with the Mazor system. One patient had bowel obstruction and acute renal injury postoperatively. There were no postoperative screw revisions or neurological deficits associated with screw placement. No complications were attributed to robot-assisted pedicle screw placement.

Discussion

In this study, we found no significant difference in the screw placement accuracy, time per screw placement, and complication rate among the 4-time periods in our initial experience, indicating no significant learning curve; further, this indicates that from the early experience of this technique, acceptable outcomes are achieved. However, the fluoroscopy time increased in the third quarter of experience (median=4 seconds). In our series, the overall complication rate was 5.5% although there were no complications attributed to mal-placement of robot-assisted pedicle screws.
‘Clinically acceptable’ screws in the literature are those deemed grade A or B (<2 mm deviation from the cortex) using the Gertzbein-Robbins method [19-21]. Of the 352 pedicle screws that were placed, 333 (94.6%) were grade A or B and 19 (5.3%) were grade C, D, or E. A more clinically relevant outcome than screw placement accuracy is screw revision. Our intraoperative revision rate was 0.3% (1 of 352) with no postoperative evaluation of the screws requiring revision. This is similar to the results from the growing body of literature that has documented high screw placement accuracy rates. Keric et al. [13] evaluated 413 patients who had Renaissance robot-guided surgery in two German centers and found a 2% misplacement rate of 3–6 mm and 1.1% deviation of >6 mm. Nine screws required revision (0.48%). Registration failure occurred in one patient because of obesity. Most deviated screws showed lateral deviation and probably resulted from ‘skiving.’ Twenty-four screw misplacements were potentially due to platform dislocation, and in six screw placements, the reason for the need of revision was unclear [13]. Kim et al. [12] performed an randomized controlled trial (RCT) of robot-assisted (n=37) versus fluoroscopy-guided (n=41) screw placement in patients undergoing posterior lumbar interbody fusion (PLIF). Although there was no difference in pedicle screw placement accuracy, robot-guided screws were less likely to breach the proximal facet joint with better convergence angles. Hyun et al. [11] reported a RCT comparing the Renaissance robot system with traditional fluoroscopy-guided pedicle screw placement in 1- or 2-level degenerative lumbar spinal fusions. Sixty patients were enrolled, and 270 screws were placed. There was 5% screw deviation in the fluoroscopy group compared to only 2.3% in the robot group; however, the placement of all the screws in the robotic group was clinically acceptable [11]. In the same study, Hyun et al. [11] compared the first 15 patients with the latter 15 patients in the robot group and found that time per screw decreased from 5.5 to 4.0 minutes (p=0.023). Fluoroscopy time per screw similarly decreased from 4.1 to 2.9 seconds; however, this was not statistically significant (p=0.117) [11].
Reporting of the learning curve in robot-assisted pedicle screw placement is sporadic and varied with outcomes not only relating to screw placement accuracy and revision, but also time per screw, fluoroscopy time per screw, and the effect of surgeon seniority on operating times being incorporated into larger studies that assess the viability of this technique.
In our study, ideal screw placement did not vary among the groups, ranging from 85.6%–95.5% (p=0.106), and our median time per screw was 7 minutes, ranging from 6.0–7.0 minutes in the four groups (p=0.605), indicating that there was no significant learning curve for this technique. However, the median fluoroscopy time varied, ranging from 2–4 seconds per screw among the four groups (p=0.021). The longest fluoroscopy times were recorded in the third quarter of our experience; this may be due to the much higher number of thoracic cases (5/18, 27.8%) in that quarter compared to that in the others (p=0.032). It is noteworthy that ideal screw placement was highest in the first quarter, probably because the surgeons were more cautious while executing a new technique and had assistive supervision.
Conflicting results have been reported for the learning curve of this technique in a large study by Schatlo et al. [18] that reported an initial learning curve that included screw placement in about 25 cases; this study reported that misplacement rates peaked between the 5th and 25th surgery in 12 of 13 surgeons in a study on 258 surgeries, and 1,265 pedicle screws that were placed with robotic assistance. Contrary to this, some other studies have shown no increase in the already high accuracy of early and late experience in robot-assisted pedicle screws [22-24]. Our study expands upon previous research because as per our initial experience, the number of patients undergoing multilevel fixations in both, the thoracic and lumbar spine and the total number of patients and screws was much larger than that in the study by Kim et al. [24] who reported 20 patients undergoing robot-assisted PLIFs. The screw placement accuracy rate for open robot-assisted pedicle screws in a pediatric population for adolescent idiopathic scoliosis was unchanged when the learning curve between the first 16 and final 16 patients was examined (9.6% versus 7.4% >2 mm deviation, p=0.573) [23]. Moreover, we measured the time per screw placement, revision rate, and fluoroscopy times to show that in other aspects of pedicle screw placement, not just accuracy there is minimal learning curve.
Fluoroscopy time decreased by 30% after the first 8 of 37 cases in a trial by Kim et al. [12] that compared robotassisted PLIF and freehand pedicle screw PLIF.
Learning curve, evaluated as time per screw placement between levels of training, was evaluated by Urakov et al. [25] in their study demonstrating that the average time per screw placement was 4.4 minutes for residents with <5 years of experience and 4.02 for residents and fellows with ≥6 years of training (p=0.61). This may indicate that the learning curve is short. Although a formal evaluation of the learning curve was not conducted, the same study noted a trend of decrease in the time needed for screw placement by residents with more experience [25].
Our results indicate that there is almost no learning curve associated with robot-assisted pedicle screw placement; however, our results may not be applicable to all neurosurgeons or orthopedic surgeons because all our authors have a subspecialty interest in spine surgery. One strength of our study is that our experience is reflective of a general spine practice, with 20 of the 73 patients undergoing multilevel fixation and 10 of the 73 patients undergoing surgery for either tumor or trauma. Moreover, our case numbers are moderate, with mal-placement and adverse event rates being low; therefore, the possible complications associated with the increasing use of robotassisted spinal surgery may have been under reported. Further large prospective studies are needed to assess the factors that affect robot screw placement accuracy and its use in different pathologies, such as a deformity with rotated pedicles or altered anatomy, for a complete evaluation of this technique.

Conclusions

Our study includes a comprehensive analysis of the learning curve associated with the adoption of robot-assisted pedicle screw placement as a new surgical technique. We evaluated multiple outcomes, such as screw placement accuracy, screw placement time, and complication rate, and found that robot-assisted screws were associated with high overall accuracy, low placement and fluoroscopy times, and low complication rates; however, there was no significant difference in these parameters from the initial experience to the practiced experience (after approximately 1 year), indicating that robot-assisted pedicle screw placement has a very short to almost no learning curve.

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Notes

Author Contributions

Study design and concept: AM, GN, BK, JK; data collection, analysis: JK, CG, SD, MA; and manuscript writing: JK, SD.

Fig. 1.
Computed tomography scans demonstrating the GertzbeinRobbins classification of screw positioning. The grades reflect the deviation of the screw from the ‘ideal’ intrapedicular trajectory. (A) Grade A is an intrapedicular screw without breach of the cortical layer of the pedicle; (B) grade B reflects a screw that breaches the cortical layer of the pedicle; however, it does not exceed it laterally by >2 mm. (C, D) Grades C and D indicate a penetration of <4 and <6 mm, respectively. (E) Grade E indicates a screw that does not pass through the pedicle or that, at any given point in its intended intrapedicular course, breaches the cortical layer of the pedicle in any direction by ≥6 mm (arrow). Reproduced from Schatlo et al. J Neurosurg Spine 2014;20:636-43, with permission from American Association of Neurological Surgeons [17].
asj-2019-0033f1.jpg
Fig. 2.
Patient flow diagram. CT, computed tomography.
asj-2019-0033f2.jpg
Fig. 3.
Scatterplot of the time per screw in minutes with increasing experience: maintenance of low time per screw from the beginning throughout the surgical experience.
asj-2019-0033f3.jpg
Table 1.
Demographic and clinical characteristics of study participants
Characteristic Total (n=73) Group 1 (n=18) Group 2 (n=18) Group 3 (n=18) Group 4 (n=19) p-valuea)
Patient demographics
 Age (yr) 66.6±11.1 69.3±8.1 65.6±8.1 67.3±11.8 64.2±14.9 0.549
 Female gender 46 (63.0) 12 (66.7) 13 (72.2) 12 (66.7) 9 (47.4) 0.414
Patient characteristics
 Diabetes 8 (11.0) 1 (5.6) 2 (11.1) 3 (16.7) 2 (10.5) 0.863
 Cardiovascular disease 18 (24.7) 7 (38.9) 2 (11.1) 4 (22.2) 5 (26.3) 0.279
 Pulmonary disease 5 (6.8) 1 (5.6) 1 (5.6) 2 (11.1) 1 (5.3) 0.934
 Osteoporosis 5 (6.8) 0 2 (11.1) 1 (5.6) 2 (10.5) 0.745
 Steroid use 5 (6.8) 2 (11.1) 2 (11.1) 1 (5.6) 0 0.469
Surgical characteristics
 Degenerative 63 (86.3) 16 (88.9) 16 (88.9) 13 (72.2) 18 (94.7)
 Neoplastic 6 (8.2) 1 (5.6) 1 (5.6) 3 (16.7) 1 (5.3) 0.588
 Trauma/fracture 4 (5.5) 1 (5.6) 1 (5.6) 2 (11.1) 0 (0.0) 0.360
 Single level/multilevel 53 (72.6)/20 (27.4) 11 (61.1)/7 (38.9) 14 (77.8)/4 (22.2) 12 (66.7)/6 (33.3) 16 (84.2)/3 (15.8) 0.407
 Thoracic/lumbar 8 (11.0) /65 (89.0) 1 (5.6)/17 (94.4) 2 (11.1) /16 (88.9) 5 (27.8)/13 (72.2) 0/19 (100.0) 0.032*
 Latera/XLIF 36 (49.3) 12 (66.7) 8 (44.4) 6 (33.3) 10 (52.6)
 Posterior/PLIF 21 (28.8) 4 (22.2) 7 (38.9) 5 (27.8) 5 (26.3) 0.588
 Anterior/ALIF/OLIF 12 (16.4) 1 (5.6) 2 (11.1) 5 (27.8) 4 (21.0) 0.194
 Pedicle screws only 4 (5.5) 1 (5.6) 1 (5.6) 2 (11.1) 0 0.411

Values are presented as mean±standard deviation or number (%).

XLIF, extreme lateral interbody fusion; PLIF, posterior lumbar interbody fusion; ALIF, anterior lumbar interbody fusion; OLIF, oblique lumbar interbody fusion.

a) From Fisher’s exact, with Freeman-Halton extension, chi-square, Mann-Whitney U-test, Kruskal-Wallis, one way analysis of variance as appropriate.

* p <0.05 (significant)

Table 2.
Screw accuracy, placement, and fluoroscopy time
Variable Total (n=73) Group 1 (n=18) Group 2 (n=18) Group 3 (n=18) Group 4 (n=19) p-valuea)
Pedicle screws placed 352 (100.0) 89 (100.0) 87 (100.0) 90 (100.0) 86 (100.0)
Ideal screw placement (%)
 Grade A screws 317 (90.1) 85 (95.5) 80 (92.0) 77 (85.6) 75 (87.2) 0.105916
 Grade B–E screws 35 (9.9) 4 (4.5) 7 (8.0) 13 (14.4) 11 (12.8)
Clinically acceptable screw placement (%)
 Grade A–B screws 333 (94.6) 86 (96.6) 83 (95.4) 86 (95.6) 78 (90.7) 0.3140
 Grade C–E screws 19 (5.4) 3 (3.4) 4 (4.6) 4 (4.4) 8 (9.3)
Distribution of screw accuracy B–E (%)
 Grade B 16 (4.5) 1 (1.1) 3 (3.4) 9 (10.0) 3 (3.5)
 Grade C 10 (2.8) 2 (2.2) 2 (2.3) 1 (1.1) 5 (5.8)
 Grade D 7 (2.0) 1 (1.1) 1 (1.1) 3 (3.3) 2 (2.3)
 Grade E 2 (0.6) 0 (0.0) 1 (1.1) 0 (0.0) 1 (1.2)
Screw placement times (interquartile range)
 Median time per screw in minutes 7.0 (3.0) 6.5 (3.0) 7.0 (4.0) 6.0 (3.0) 6.0 (4.0) 0.605
 Median fluoroscopy time per screw in seconds 3.0 (2.0) 2.0 (2.0) 3.0 (1.0) 4.0 (9.0) 3.0 (2.0) 0.021*

Values are presented as number (%), unless otherwise stated.

a) From Fisher’s exact, with Freeman-Halton extension, chi-square, Mann-Whitney U-test, Kruskal-Wallis, one way analysis of variance as appropriate.

* p <0.05 (significant).

Table 3.
Screw misplacement, adverse events, and complications
Variable Total (n=73) Group 1 (n=18) Group 2 (n=18) Group 3 (n=18) Group 4 (n=19) p-valuea)
Pedicle screws placed 352 (100.0) 89 (100.0) 87 (100.0) 90 (100.0) 86 (100.0)
Intraoperative screw revision 1 (0.3) 0 0 0 1 (1.2)
Misplaced screws 0 0 0 0 0
Complications (%) 4 (5.5) 0 1 (5.6) 3 (16.7) 0 0.093

Values are presented as number (%), unless otherwise stated.

a) From Fisher’s exact, with Freeman-Halton extension, chi-square, Mann-Whitney U-test, Kruskal-Wallis, one way analysis of variance as appropriate.

* p <0.05 (significant).

References

1. Babu R, Park JG, Mehta AI, et al. Comparison of superior-level facet joint violations during open and percutaneous pedicle screw placement. Neurosurgery 2012 71:962–70.
crossref pmid pmc pdf
2. Chen Z, Zhao J, Xu H, Liu A, Yuan J, Wang C. Technical factors related to the incidence of adjacent superior segment facet joint violation after transpedicular instrumentation in the lumbar spine. Eur Spine J 2008 17:1476–80.
crossref pmid pmc pdf
3. Gejo R, Matsui H, Kawaguchi Y, Ishihara H, Tsuji H. Serial changes in trunk muscle performance after posterior lumbar surgery. Spine (Phila Pa 1976) 1999 24:1023–8.
crossref pmid
4. Kim HJ, Chun HJ, Kang KT, et al. The biomechanical effect of pedicle screws’ insertion angle and position on the superior adjacent segment in 1 segment lumbar fusion. Spine (Phila Pa 1976) 2012 37:1637–44.
crossref pmid
5. Knox JB, Dai JM 3rd, Orchowski JR. Superior segment facet joint violation and cortical violation after minimally invasive pedicle screw placement. Spine J 2011 11:213–7.
crossref pmid
6. Lee KH, Yue WM, Yeo W, Soeharno H, Tan SB. Clinical and radiological outcomes of open versus minimally invasive transforaminal lumbar interbody fusion. Eur Spine J 2012 21:2265–70.
crossref pmid pmc pdf
7. Moshirfar A, Jenis LG, Spector LR, et al. Computed tomography evaluation of superior-segment facetjoint violation after pedicle instrumentation of the lumbar spine with a midline surgical approach. Spine (Phila Pa 1976) 2006 31:2624–9.
crossref pmid
8. Park Y, Ha JW. Comparison of one-level posterior lumbar interbody fusion performed with a minimally invasive approach or a traditional open approach. Spine (Phila Pa 1976) 2007 32:537–4.
crossref pmid
9. Mobbs RJ, Sivabalan P, Li J. Technique, challenges and indications for percutaneous pedicle screw fixation. J Clin Neurosci 2011 18:741–9.
crossref pmid
10. Sclafani JA, Kim CW. Complications associated with the initial learning curve of minimally invasive spine surgery: a systematic review. Clin Orthop Relat Res 2014 472:1711–7.
crossref pmid pmc
11. Hyun SJ, Kim KJ, Jahng TA, Kim HJ. Minimally invasive robotic versus open fluoroscopic-guided spinal instrumented fusions: a randomized controlled trial. Spine (Phila Pa 1976) 2017 42:353–8.
crossref pmid
12. Kim HJ, Jung WI, Chang BS, Lee CK, Kang KT, Yeom JS. A prospective, randomized, controlled trial of robot-assisted vs freehand pedicle screw fixation in spine surgery. Int J Med Robot 2017 13.
crossref
13. Keric N, Doenitz C, Haj A, et al. Evaluation of robotguided minimally invasive implantation of 2067 pedicle screws. Neurosurg Focus 2017 42:E11.
crossref
14. Schroder ML, Staartjes VE. Revisions for screw malposition and clinical outcomes after robot-guided lumbar fusion for spondylolisthesis. Neurosurg Focus 2017 42:E12.
crossref pmc
15. Devito DP, Kaplan L, Dietl R, et al. Clinical acceptance and accuracy assessment of spinal implants guided with SpineAssist surgical robot: retrospective study. Spine (Phila Pa 1976) 2010 35:2109–15.
crossref pmid
16. Hu X, Ohnmeiss DD, Lieberman IH. Robotic-assisted pedicle screw placement: lessons learned from the first 102 patients. Eur Spine J 2013 22:661–6.
crossref pmid pdf
17. Schatlo B, Molliqaj G, Cuvinciuc V, Kotowski M, Schaller K, Tessitore E. Safety and accuracy of robotassisted versus fluoroscopy-guided pedicle screw insertion for degenerative diseases of the lumbar spine: a matched cohort comparison. J Neurosurg Spine 2014 20:636–43.
crossref pmid
18. Schatlo B, Martinez R, Alaid A, et al. Unskilled unawareness and the learning curve in robotic spine surgery. Acta Neurochir (Wien) 2015 157:1819–23.
crossref pmid pdf
19. Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo. Spine (Phila Pa 1976) 1990 15:11–4.
crossref pmid
20. Chenin L, Capel C, Fichten A, Peltier J, Lefranc M. Evaluation of screw placement accuracy in circumferential lumbar arthrodesis using robotic assistance and intraoperative flat-panel computed tomography. World Neurosurg 2017 105:86–94.
crossref pmid
21. Fan Y, Du JP, Liu JJ, et al. Accuracy of pedicle screw placement comparing robot-assisted technology and the free-hand with fluoroscopy-guided method in spine surgery: an updated meta-analysis. Medicine (Baltimore) 2018 97:e10970.
crossref pmid pmc
22. Ringel F, Stüer C, Reinke A, et al. Accuracy of robotassisted placement of lumbar and sacral pedicle screws: a prospective randomized comparison to conventional freehand screw implantation. Spine (Phila Pa 1976) 2012 37:E496–501.
crossref pmid
23. Macke JJ, Woo R, Varich L. Accuracy of robot-assisted pedicle screw placement for adolescent idiopathic scoliosis in the pediatric population. J Robot Surg 2016 10:145–50.
crossref pmid pdf
24. Kim HJ, Lee SH, Chang BS, et al. Monitoring the quality of robot-assisted pedicle screw fixation in the lumbar spine by using a cumulative summation test. Spine (Phila Pa 1976) 2015 40:87–94.
crossref pmid
25. Urakov TM, Chang KH, Burks SS, Wang MY. Initial academic experience and learning curve with robotic spine instrumentation. Neurosurg Focus 2017 42:E4.
crossref


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
Department of Orthopedic Surgery, Asan Medical Center, University of Ulsan College of Medicine
88, Olympic-ro 43-gil, Songpa-gu, Seoul 05505, Korea
Tel: +82-2-3010-3530    Fax: +82-2-3010-8555    E-mail: asianspinejournal@gmail.com                
Korean Society of Spine Surgery
27, Dongguk-ro, Ilsandong-gu, Goyang-si 10326, Korea
Tel: +82-31-966-3413    Fax: +82-2-831-3414    E-mail: office@spine.or.kr                

Copyright © 2024 by Korean Society of Spine Surgery.

Developed in M2PI

Close layer
prev next