Assessing the stiffness of spinal fusion in animal models

HSS J. 2006 Feb;2(1):12-8. doi: 10.1007/s11420-005-5123-7.

Abstract

The clinical goal of spinal fusion is to reduce motion and the associated pain. Therefore, measuring motion under loading is critical. The purpose of this study was to validate four-point bending as a means to mechanically evaluate simulated fusions in dog and rabbit spines. We hypothesized that this method would be more sensitive than manual palpation and would be able to distinguish unilateral vs bilateral fusion. Spines from four mixed breed dogs and four New Zealand white rabbits were used to simulate posterolateral fusion with polymethyl methacrylate as the fusion mass. We performed manual palpation and nondestructive mechanical testing in four-point bending in four planes of motion: flexion, extension, and right and left bending. This testing protocol was used for each specimen in three fusion modes: intact, unilateral, and bilateral fusion. Under manual palpation, all intact spines were rated as not fused, and all unilateral and bilateral simulated fusions were rated as fused. In four-point bending, dog spines were significantly stiffer after unilateral fusion compared with intact in all directions. Additionally, rabbit spines were stiffer in flexion and left bending after unilateral fusion. All specimens exhibited significant differences between intact and bilateral fusion except the rabbit in extension. For unilateral vs bilateral fusion, significant differences were present for right bending in the dog model and for flexion in the rabbit. Unilateral fusion can provide enough stability to constitute a fused grade by manual palpation but may not provide structural stiffness comparable to bilateral fusion.