TY - JOUR T1 - Stability and Load Sharing Characteristics of a Posterior Dynamic Stabilization Device JF - International Journal of Spine Surgery JO - Int J Spine Surg DO - 10.14444/2009 VL - 9 SP - 9 AU - Daniel J. Cook AU - Matthew S. Yeager AU - Shankar S. Thampi AU - Donald M. Whiting AU - Boyle C. Cheng Y1 - 2015/01/01 UR - http://ijssurgery.com//content/9/9.abstract N2 - Background Lumbar interbody fusion is a common treatment for a variety of spinal pathologies. It has been hypothesized that insufficient mechanical loading of the interbody graft can prevent proper fusion of the joint. The purpose of this study was to evaluate the mechanical stability and anterior column loading sharing characteristics of a posterior dynamic system compared to titanium rods in an anterior lumbar interbody fusion (ALIF) model.Methods Range of motion, interpedicular kinematics and interbody graft loading were measured in human cadaveric lumbar segments tested under a pure moment flexibility testing protocol.Results Both systems provided significant fixation compared to the intact condition and to an interbody spacer alone in flexion extension and lateral bending. No significant differences in fixation were detected between the devices. A significant decrease in graft loading was detected in flexion for the titanium rod treatment compared to spacer alone. No significant differences in graft loading were detected between the spacer alone and posterior dynamic system or between the posterior dynamic system and the titanium rod.Conclusions The results of this study indicate that the posterior dynamic system provides similar fixation compared to that of a titanium rod, however, studies designed to evaluate the efficacy of fixation in a cadaver model may not be sufficiently powered to establish differences in load sharing using the techniques described here. ER -