Table 3

Hardware failure (21 patients)—2-y follow-up.

Mode of Failure n WomenMen
Broken screw211
Screw loosening532
Symptomatic hardware prominence642
Set Screw loosening4a 22
Rod fracturesc 6b 42
Total 23 14 9
  • a Three of the set screw loosenings were asymptomatic. No revision required.

  • b Five rod fractures were asymptomatic with solid fusion on computed tomography scan. No revision.

  • c One male patient had symptomatic T12- L1 and L5-S1 nonunion with bilateral distal rod fractures, iliac set screw loosening, and proximal T12 screw loosening.