Elsevier

The Spine Journal

Volume 9, Issue 3, March 2009, Pages 263-267
The Spine Journal

Technical Review
Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation

https://doi.org/10.1016/j.spinee.2008.08.002Get rights and content

Abstract

Background context

Polyetheretherketone (PEEK) has been increasingly used as a biomaterial for spinal implants. PEEK lumbar fusion rods have recently become available for use in posterior lumbar fusion procedures.

Purpose

To compare Polyetheretherketone Rod System to traditional titanium rod fixation in a cadaveric model and provide mechanical test data for the PEEK system.

Study design

Biomechanical testing.

Methods

Cadaveric biomechanical testing was conducted to compare Expedium 5.5 mm PEEK rods to titanium rods of equivalent diameter. Biomaterials testing was performed to determine static and dynamic performance of Expedium 5.5 mm PEEK rods with 6% BaSo4 in compressive bending and torsion.

Results

Cadaveric testing demonstrated that PEEK rods can significantly reduce the range of motion of a destabilized segment. The testing showed no significant difference in the stability provided by PEEK and titanium rods in posterolateral fusion (PLF) or posterior lumbar interbody fusion (PLIF) constructs. PEEK static compressive bending tests showed 67 degrees displacement without fracture of the rod. Torsion testing showed 30 degrees of rotation without yield or plastic deformation. Dynamic compression testing revealed two fatigue runouts at 23 degrees.

Conclusions

PEEK rods provide comparable stability to titanium rods of equivalent diameter in cadaveric testing. Mechanical testing suggests PEEK rods can withstand far beyond the angular displacements suggested by cadaveric testing and that of normal physiologic range of motion. Potential advantages to PEEK rods include better anterior column load sharing, reduced stress at bone-to-screw interface, and reduced computed tomography and magnetic resonance imaging scatter and artifact.

Introduction

The use of lumbar pedicle screw instrumentation to augment interbody and posterolateral fusion (PLF) rates has been well established [1], [2], [3], [4], [5]. Traditional instrumentation consisting of stainless steel constructs eventually evolved into titanium implants as a result of closer biomechanical properties to bone, improved biocompatibility, and reduced magnetic resonance imaging (MRI) artifact. Titanium offers sufficient strength under physiologic loads and construct stiffness closer to the modulus of elasticity of cortical bone, thus improving the stress shielding characteristics of lumbar pedicle instrumentation [3], [4], [6], [7], [8].

Recent advancements in posterior spinal fixation has centered on the concept of dynamic stabilization [9], [10], [11], [12], [13]. Often used as a nonfusion alternative, dynamic stabilization has a theoretical advantage to traditional stiff fixation of minimizing adjacent segment disc degeneration. When applied to fusion, a more dynamic stabilization may provide additional load sharing onto the anterior column and reduce stresses at the bone screw interface. In the common case of an undersized graft or graft subsidence, these advantages may be amplified as a system of lower stiffness may flex to appropriately load the interbody device. However, inherent to reduced fixation stability is the risk of pseudarthrosis. The ideal fixation system would maximize fusion rates by providing sufficient stability without excessive rigidity to allow for bone graft loading, while maintaining natural posturing and alignment to minimize adjacent level stress.

Polyetheretherketone (PEEK) has been increasingly used as a biomaterial for trauma, orthopedic, and spinal implants [7], [14]. PEEK is a thermoplastic polymer whose chemical structure maintains stability at temperatures exceeding 300°C, resists chemical and radiation damage, exhibits greater strength per mass than many metals, and offers compatibility with many reinforcing agents. PEEK as a biomaterial is fully biocompatible, with numerous studies documenting minimal systemic, intracutaneous, and intramuscular toxicity [7], [14], [15]. Furthermore, PEEK is considered to be relatively inert biologically with no evidence of inflammatory reaction to wear debris. Rivard et al. concluded after an in vivo biocompatibility study on New Zealand White rabbits that PEEK polymer is harmless to the spinal cord and it may be used safely as a component of spinal implants [16].

The purpose of this study was to compare the biomechanical performance of PEEK rods to titanium rods in a cadaveric study when used as posterior spinal instrumentation.

In addition, mechanical testing is provided that characterizes the properties of 5.5 mm PEEK lumbar rods.

Section snippets

Cadaveric testing

Four fresh human cadaveric lumbar spines from L1 to S1 were harvested and used for this experiment after screening for abnormal anatomy using anteroposterior and lateral fluoroscopy. The specimens were stored in double plastic bags at −20°C until preparation and testing. Both L1 and sacrum were rigidly fixed and potted into custom cups using wood screws and a urethane compound (SmoothCast [Smooth-On Inc., Easton, PA]) such that the specimen was in neutral posture with the L3–L4 disc oriented

Cadaveric testing

Results demonstrated that the PLIF and PLF PEEK constructs significantly reduced the range of motion (ROM) as compared with both the intact and destabilized conditions (p<.05). Results demonstrated that there was no significant difference in the stability provided by a 5.5 mm PEEK construct and a 5.5 mm Titanium construct (Table 2). After destabilization, there was no significant difference (p<.05) in the percent reduction in ROM provided by Titanium PLF, PEEK PLF, Titanium PLIF, or PEEK PLIF for

Discussion

PEEK rods have recently become available for use in posterior lumbar fusion procedures. To the best of our knowledge, there have been no published reports of the biomechanical properties of these devices. It has been suggested that stiffness of metallic spine implants/constructs may far exceed the requirements for successful fusion and predispose secondary adjacent level disc degeneration and failure [12].

The results from our cadaveric testing show that the PEEK rod construct significantly

Conclusion

PEEK rods offer an alternative to traditional Titanium rod fixation. Biocompatibility studies, biomechanical evaluation, and cadaveric testing demonstrate that PEEK rods can provide stability under normal physiologic conditions. In addition, several notable benefits of PEEK in comparison to Titanium include increased anterior column load sharing, reduced stress at bone-to-screw interface, and reduced computed tomography and MRI scatter/artifact.

Further study is needed to evaluate the clinical

References (22)

  • M. Wedemeyer et al.

    Titanium versus stainless steel for anterior spinal fusions: an analysis of rod stress as a predictor of rod breakage during physiologic loading in a bovine model

    Spine

    (2007)
  • Cited by (146)

    • Advances in Implant Technologies for Spine Surgery

      2024, Neurosurgery Clinics of North America
    • Cold spray technology: a perspective of nature-inspired feature processing and biomanufacturing by a heatless additive method using nanopowders

      2022, Advances in Additive Manufacturing: Artificial Intelligence, Nature-Inspired, and Biomanufacturing
    • Evolution of materials for implants in metastatic spine disease till date – Have we found an ideal material?: Evolution of implant materials in metastatic spine disease

      2021, Radiotherapy and Oncology
      Citation Excerpt :

      PEEK, a biocompatible polymer, has been used in spinal implants such as fusion cages since the 1990s [46]. It is bioinert and has a modulus of elasticity of 3.6 GPa which is lower than that of cortical bone, thus causing minimal stress shielding and better load sharing [47]. Additionally, PEEK is radiolucent and non-metallic, producing minimal artifacts on CT and MRI [45].

    View all citing articles on Scopus

    FDA device/drug status: not applicable.

    The following authors acknowledge a financial relationship: ARV and TA: Royalties; HS and BZ: Other office in the company, which may indirectly relate to the subject of this research.

    View full text