Anatomically correct reduction and fixation of a Tile C-1 type unilateral sacroiliac disruption using a rod and pedicle screw system between the S1 vertebra and the ilium: experimental and clinical case report

J Orthop Sci. 2002;7(5):581-6. doi: 10.1007/s007760200104.

Abstract

We have developed a new surgical technique for the treatment of Tile C-1 type sacroiliac disruption. We tried this procedure first in a cadaveric specimen and then applied it to a clinical case. We used the Texas Scottish Rite Hospital (TSRH) rod and pedicle screw system to insert one screw into the S1 vertebra without using an image intensifier and the other screw into the bone marrow of the ilium from the posterosuperior iliac spine. A straight rod was connected between the two screws by using a manipulator to attempt to reduce and fix the sacroiliac disruption. The combined pubic symphysis diastasis could be simultaneously reduced and fixed by using a plate through another incision, resulting in anatomically correct reconstruction of the pelvic ring. In this procedure, the alignment of the sacroiliac joint can be reversibly and directly changed during reduction and fixation. The sacroiliac joint can be strongly fixed because the screws can be freely inserted into the intact portion of the pelvis and the adjacent lumbar spine, if necessary. Good reduction is obtained because direct compression force is applied to the fracture site. The posterior and anterior procedures can be simultaneously performed under the same lateral position.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Bone Screws
  • Female
  • Hip Joint
  • Humans
  • Internal Fixators*
  • Joint Instability / surgery*
  • Orthopedic Procedures
  • Pelvic Bones / injuries*
  • Sacroiliac Joint / surgery*