Release of cobalt and chromium ions into the serum following implantation of the metal-on-metal Maverick-type artificial lumbar disc (Medtronic Sofamor Danek)

Spine (Phila Pa 1976). 2007 Feb 1;32(3):348-52. doi: 10.1097/01.brs.0000253599.89694.c0.

Abstract

Study design: Cross-sectional study of 10 patients to measure the serum levels of cobalt and chromium after TDA.

Objectives: To investigate the release of cobalt and chromium ions into the serum following implantation of the metal-on-metal Maverick-type artificial lumbar disc.

Summary of background data: In total hip endoprosthetics and consequently for TDA (total disc arthroplasty), metal-on-metal combinations are used with the aim of reducing wear debris. In metal-on-metal TDA the release of metal ions has until now been secondary to the main discussion.

Materials and methods: We investigated the serum cobalt and chromium concentration following implantation of 15 Maverick TDAs (monosegmental L5-S1, n = 5; bisegmental L4-L5 and L5-S1, n = 5; average age, 36.5 years). Five healthy subjects (no metal implants) acted as a control group. The measurements of the metals were carried out using the HITACHI Z-8200 AAS polarized Zeeman atomic absorption spectrometer after an average of 14.8 months.

Results: The concentrations of cobalt and chromium ions in the serum amounted on average to 4.75 microg/L (SD, 2.71) for cobalt and 1.10 microg/L (SD, 1.24) for chromium. Compared with control group, both the chromium and cobalt levels in the serum showed significant increases (Mann-Whitney U test, P = 0.0120). At follow-up,the Oswestry Disability Score was on average significantly decreased by 24.4 points (L5-S1) (t test, P < 0.05) and by 26.8 points (L4-S1) (t test, P < 0.05). The improved clinical situation is also represented by a significant decrease of the Visual Analog Pain Scale of 42.2 points after the follow-up (t test, P < 0.05).

Conclusion: Significant systemic release of Cr/Co was proven in the serum compared with the control group. The concentrations of Cr/Co measured in the serum are similar in terms of their level to the values measured in THA metal-on-metal combinations or exceed these values given in the literature. Long-term implication of this metal exposure is unknown and should be studied further.

Publication types

  • Comparative Study

MeSH terms

  • Adolescent
  • Adult
  • Chromium / blood*
  • Cobalt / blood*
  • Cross-Sectional Studies
  • Female
  • Follow-Up Studies
  • Humans
  • Intervertebral Disc / metabolism
  • Intervertebral Disc / surgery*
  • Lumbar Vertebrae / metabolism
  • Lumbar Vertebrae / surgery*
  • Male
  • Middle Aged
  • Prosthesis Design / adverse effects
  • Prosthesis Design / instrumentation
  • Prosthesis Implantation / adverse effects*
  • Prosthesis Implantation / instrumentation

Substances

  • Chromium
  • Cobalt